6. This step is to adjust Span. The Wizard will output +10 V on the
channel you selected. The steps of span calibration are quite similar
to those of the offset calibration (Step 5). Follow the instructions of
the Calibration Wizard. If value fits the condition, press the [Next]
button to continue.
Figure D.5: The Adjustment Process of Span Calibration
Step 7: After span calibration, you have successfully calibrated your PCI-
1724U device. Please press [Finish] to finish it.
Figure D.6: Finish Calibration